Chinese Journal of Breast Disease(Electronic Edition) ›› 2009, Vol. 03 ›› Issue (04): 398-405. doi: 10.3877/cma.j.issn.1674-0807.2009.04.005

• Clinical Research • Previous Articles     Next Articles

Clinical value of diffused optical tomography with ultrasound in the diagnosis of breast cancer

Yue LI1, Ying ZHAO1, Bin ZHANG1, Jie GE1, Lian-sheng NING1, Xuchen CAO1, Zhi-hao YU1, Xin WANG,1()   

  1. 1.First Department of Breast Tumor,Cancer Hospital and Institute,Tianjin Medical University,Tianjin 300060,China
  • Received:2009-04-03 Online:2009-08-01 Published:2024-12-12
  • Contact: Xin WANG

Abstract:

Objective

To study the value of diffused optical breast imaging with ultrasound(OPTIMUS®,OPTIMUS system)in the diagnosis of breast tumors.

Methods

The images of 136 breast tumors by OPTIMUS system were reviewed and compared with pathological results.Meanwhile ultrasound and X-ray examinations were performed.Consistency checking was used to evaluate the consistency of pathological, ultrasound and X-ray examinations. The receiver operating characteristic(ROC)curve was done to calculate the area under the curve to evaluate the accuracy of OPTIMUS system and the clinical value of OPTIMUS system in the diagnosis of breast cancer.

Results

In the 136 cases,pathological diagnosis showed 50 cases were benign and 86 cases malignant.OPTIMUS system diagnosis showed 34 cases were benign and 102 cases malignant;among the 102 cases,16 were overdiagnosed as cancer.Bases on the pathological results,the consistency checking Kappa value was 0.536 and 0.386 for ultrasound and X-ray examinations respectively,and 0.729,0.616,and 0.649 for OPTIMUS system,ultrasound and Xray examinations respectively.The area under ROC for OPTIMUS system,ultrasound and X-ray examinations was 0.840(95%CI 0.758-0.922),0.817(95%CI 0.739-0.895)and 0.837(95%CI 0.763-0.910)respectively.

Conclusion

OPTIMUS system is a non-invasive and highly effective diagnostic tool for breast disease and it could be used as assistant diagnostic tool for breast tumors.

Key words: Diffused optical tomography with ultrasound, Breast tumor, Ultrasonography, Mammography