Set-up errors in helical tomotherapy for breast cancer patients

Fuhai Zhu1, Xiaolong Hu1, Hefei Liu1, Weizhang Wu1, Yingjie Wang1,(), Tingyi Xia1,()   

  1. 1.Department of Radiation Oncology, Air Force General Hospital, Beijing 100142, China
  • Received:2015-04-23 Online:2015-06-01 Published:2024-12-13
  • Contact: Yingjie Wang, Tingyi Xia

Abstract:

Objective

To analyze the set-up errors of breast cancer patients based on helical tomotherapy treatment with mega-voltage CT (MVCT), and measure the margins of planning target volume(PTV).

Methods

We enrolled fourteen breast cancer patients treated with helical tomotherapy in Department of Radiation Oncology, Air Force General Hospital from August 2011 to November 2014.MVCT scanning was performed before each treatment and registered with the kilo-voltage CT (kVCT) image.The set-up errors of the transverse profile rotation (Roll), left-right (X), anterior-posterior (Y) and superior-inferior (Z) positions were recorded.The PTV margin was calculated based on the formula M=2.5∑+0.7δ-3.

Results

Three hundred and fourteen MVCT images were received for all patients.The set-up error frequency within 5 mm in the X, Y and Z directions was 89.81%(282/314), 82.17%(258/314) and 97.13%(305/314) respectively,and within 2° in the Roll direction was 91.08%(286/314).The average set-up errors were (4.26±3.70) mm,(3.10±2.80) mm, (2.02±1.50) mm and (0.70±0.40)° in the X, Y, Z and Roll directions respectively.The PTV margins in X, Y, and Z directions were 4.56,3.44 and 2.79 mm.

Conclusions

The breast cancer patients' set-up errors are acceptable based on MVCT of helical tomotherapy, and it can provide a certain quality assurance with the PTV margins of 5,3 and 3 mm in X, Y and Z directions respectively.

Key words: Breast neoplasms, Radiotherapy, Tomography, spiral computed